PART NUMBER: 198705

 

Easy Pulse 5 - Oxygen Conserving Regulator

The Oxygen Conserver is designed to be used with high-pressure oxygen systems. It consists of a cylinder connection, cylinder contents gauge, high to low pressure regulator, orifice plate and a conserving demand module. The regulator reduces the high pressure of the cylinder to the working pressure of the orifice plate. The orifice place uses calibrated orifices to deliver a selected flow to the conserving demand module. The conserving demand module controls the pulse size and timing of the patient. It supplies a pulse of oxygen at the beginning of each breath. This reduces the oxygen demand on the system and limits the drying of the airways. The flow is determined by setting the flow control knob to the prescribed flow. The oxygen is supplied to the patient through the cannula.

 

Specifications:

Inlet Pressure Range:             300-30000 PSI (2068 – 20684 kPa)

Internal Regulated Pressure: 50 PSI (345 kPa)

Pressure Gauge Accuracy:    3 – 2 – 3% of Full Scale

Dimensions:                           

-          Weight:                       279g (9.9oz)

-          Overall Length:            12.07cm (4.75in)

-          Width:                         6.67cm (2.63in)

-          Height:                        6.34cm (2.50in)

Pulse Settings:                         1, 2, 3, 4 & 5 LPM Equivalents

Flow Settings:                         2 LPM Continuous

Accuracy:                               Within +/- 15% of the nominal bolus value (at each breath rate)

                                                Continuous Setting:      2 LPM ± 10%

Savings Ratio:                                    Up to 5.7:1

Trigger Method:                     Inspiratory Effort (Negative Pressure from Patient Inhalation)

Breathing Frequency:                        Up to 35 BPM

Cannula Requirement:           Maximum 2.1m (7.1ft) long standard adult single lumen nasal cannula

Operating Conditions:           

-          Temperature:   1.7°C to 40.6°C (35°F to 105°F)

-          Altitude:           Sea Level to 0 to 3,048m (10,000ft)

Storage Conditions:

-          Temperature:               -40°C to 60°C (-40F° to 140°F)

-          Maximum Humidity:   95% Non-condensing

Oxygen Cylinder Connection:           CGA 870 Valve

Ignition and Fault Tolerance:            Meets ASTM G175-03

 

 



 

 